BOD Instrumentation Product Description
OxiTop IS 6, IS 12
Measurement using OxiTop® Biological Oxygen Demand (BOD)
Instrumentation is based on pressure measurement in a
closed system: microorganisms in the sample consume the
oxygen and form CO2. This is absorbed by
NaOH, creating a vacuum which can be read directly as a
measured value in mg/l BOD.
The sample volume being tested regulates the amount
of oxygen available for a complete BOD measurement. BOD
measurement ranges of up to 4,000 mg/l can be measured
using different sample volumes.
The OxiTop® BOD instrumentation heads (green and
yellow for differentiation of inflow/outflow) have an
AutoTemp function: if the sample temperature is too
cold, the start of BOD measurement is automatically
delayed (by at least 1 hour) until a constant
temperature has been reached.
Apart from the automatic storage of 5 measured values
(1 value per day), further measured BOD values can be
read at all times during or after the period of 5 days,
which permits the tracking of check values or
measurements over longer periods. The OxiTop BOD
Instrumentation measurement system also offers simple
operation, improved controllability, and non-toxicity.
OxiTop Control 6, Control 12
OxiTop® Control Biological Oxygen Demand (BOD)
Instrumentation is an advanced version of the successful
OxiTop® BOD Instrumentation system that uses
software-controlled functions and an infrared interface
to communicate with a handy controller, the OC 100.
This BOD instrumentation system enables the simultaneous
and grouped start, management, storage and tracking of
100 BOD measuring heads via the controller and allows
them to be tracked on a large display with graphic
evaluation. BOD measurement data can be transferred to
the PC for evaluation and documentation via the AK-540/B
cable (order no. 902842) and the communication program
Achat OC (order no. 208990).
The Biological Oxygen Demand (BOD) data can be called
up at any time, even during sampling, thus enabling
checking of the samples for errors. The display of the
progress curve allows immediate detection of
irregularities and interferences, such as having the BOD
value set too high for the volume used or undesired
nitrification. BOD measurement corrections can thus be
made at an early stage. OxiTop® Control BOD
instrumentation offers you simple operation, improved
controllability, non-toxicity, and measuring ranges of
up to 4,000 mg/l BOD.
Biological Oxygen Demand Measurements according to
EN 1899-1 and EN 1899-2 and for self-checks
Biological oxygen demand (BOD) determination is still
one of the most important parameters in water resource
management. BOD measurements can be used to evaluate
the impact of biodegradable substances in waters and
wastewater by measureing the quality of water and
treatment results in wastewater. In addition, BOD
instrumentation measurments are used in the planning and
design of wastewater treatment facilities. The OxiTop®
BOD measurement instrumentation system offers a unique,
modular and mercury-free way to make BOD measurements.
It is not only suitable for BOD determination, but also
for measuring biodegradability and depletion.
In routine use, BOD determination is used to check
the wastewater in the inflow and discharge of wastewater
treatment plants. Depending on the measurement site and
type of wastewater the BOD value can lie between a few
mg/l and several thousand mg/l. Several BOD testing
methods are available for carrying out the measurement.
Various systems for BOD determination methods
In “dilution BOD” the oxygen content of a sample is
measured with an oxygen sensor before and after an
incubation period of 5 days. The difference between the
measurements is the BOD5 value; this is the official EPA
BOD measurement method. The Oxi 1970i Portable
Dissolved Oxygen Meters can be used for dilution BOD
measurments.
In “BOD self-checks” with the respirometer the
reduction in oxygen causes a definite pressure
difference which can be measured by a pressure sensor.
This BOD measurement method is very easy and practical
to carry out.
Although both methods are very different, the BOD
measurements correlate for discharge analysis in
municipal wastewater treatment facilities. Both BOD
measurement methods require that the samples to be kept
at 68 °F (20 °C) for 5 days. WTW offers a temperature
controlled box for use with the Oxitop® BOD
instrumentation system.
BOD Instrumentation Specifications
OxiTop IS
Measuring principle: Respirometric (manometric)
Display: 2-digit 7-segment LED, 10 mm
Measured parameter: Digits
Measuring range: 0 to 40 digits (+10 digits for
overranging)
Accuracy: + 0.03 psi (+ 2 hPa) or
+ 1 digit
Admissible measuring temperature: 68°F ± 1.8°F
(20°C ± 1°C)
Admissible sample temperature (when being filled):
59-68°F (15-20°C)
Power supply: Lithium batteries (260 mAh), 2 x
type CR 2430 (3V)
Power consumption: Max. 25 mA (during
measurement)
Battery run time: ~ 3 years with normal use
Protection class: 3 IEC 1010
Protection system: IP 54 IEC 529
EMC
Emissions: EN 50081-1, FCC Class A
Immunity: EN 50082-2, NAMUR recommendation
Climate class: 2, VDI/VDE 3540
Ambient temperature:
Storage: -13°F to +149°F (-25°C to +65°C)
Operation: 68°F ± 1.8 F (20°C ± 1 C)
Test mark: CE
Dimensions: H: 2.7in, dia. 2.8in (69 mm, dia. 70
mm)
Weight: Approx. 3 oz. (85 g)
OxiTop -C
Measuring principle: Respirometric (manometric)
Measured of: BODn (oxygen drift
according to DIN 38409 T 52)
Pressure range: 7.25 to 19.5 psi (500 to 1350
hPa(mbar))
Accuracy: + 1% of measured value, +
7% of measuring range
Resolution: 0.015 psi (1 hPa)
Display: LED pilot lamps
Power supply: Lithium batteries (260 mAh), 2 x
type CR 2430 (3V)
Battery run time: ~ 2 years with normal use
Protection class: 3 IEC 1010, EN61010 part 1
Protection type: IP 54 DIN 40050
Admissible measuring temperature: 68°F ± 1.8 F
(20°C ± 1 C)
Admissible sample temperature (when being filled):
59-68°F (15-20°C)
EMC
Emissions: EN 50081-1, FCC Class A
Immunity: EN 50082-2
Climate class: 2, VDI/VDE 3540
Ambient temperature:
Storage: -13°F to +149°F (-25°C to +65°C)
Operation: 41°F to 122°F (5°C to +50°C)
Test mark: CE, UL, CUL(UL/CSA)
Dimensions: H: 2.8in, dia. 2.8in (70 mm, dia. 70
mm)
Weight: Approx. 3.4 oz. (95 g)
OC 100 BOD Instrumentation Controller
Data sets per measurement: 180 to 360 (depending
on duration)
Measurement period: 0.5 hr to 99 days
Power supply: 3 AA batteries; alkaline 1.5V
Interface: IR (infrared); RS 232 for
communications with PC
Ambient temperature:
Storage: -13°F to +149°F (-25°C to +65°C)
Operation: 41°F to 104°F (5°C to +40°C)
Dimensions (HxWxD): 1.7x3.9x7.9 in (45x100x200
mm)
Weight: Approx. 13.8 oz. (390 g)
BOD Instrumentation Magnetic Stirrer
Power supply:
Input: 230 VAC (+10%/-15%) 50/60 Hz /24 VA
Output: 18 VAC / 1 A
Power consumption: 24 VA max.
Speed of the stirring rods: Program-controlled
180 to 450 min-1
Instrument protection class: 3, IEC 1010, IP30
EMC
Emissions: EN50081-1/FCC class A
Immunity: EN50082-1, EN50082-2, Namur
Climate class: 2, VDI/VDE 3540
Ambient temperature:
Storage: -13°F to +149°F (-25°C to +65°C)
Operation: 41°F to +104°F (+5°C to +40°C)
Relative humidity:
Yearly mean: < 75 %
30 days/year: < 95 %
Remaining days: < 85 %
Test mark: CE, UL, CSA
Dimensions (LxWxH):
IS 6: 10.5x7.25x2.75 in (265x181x67 mm)
IS 12: 13.75x10.5x2.75 in (350x266x67 mm)
Weight (inclusive line adaptor):
IS 6: 3.5 lbs (1.6 kg)
IS 12: 7 lbs (3.2 kg)
BOD Instrumentation Thermostat Box
Temperature control: 68°F +0.9°F (20°C
+0.5°C)
Ambient temperature:
Storage: -13°F to +122°F (-25°C to +50°C)
Operation: 50°F to 89.6°F (10°C to 32°C)
Power consumption: 200W
Dimensions (HxWxD): 14.76x16.73x23.62 in
(375x425x600 mm)
Weight: Approx. 66.1 lbs. (30 kg)
BOD Instrumentation Options
208214 OxiTop® IS 6
BOD measuring instrument for self-check with 6
measuring units, 115 VAC, 50/60 Hz power plug, and
accessories. |
|
|
208215 OxiTop® IS 12
BOD measuring instrument for self-check with 12
measuring units, 115 VAC, 50/60 Hz power plug, and
accessories. |
|
|
208202 OxiTop® Control 6
BOD measuring instrument for self-check, with
controller OC100, 6 measuring units with IR
interface, 115 VAC, 50/60 Hz, and accessories. For
PC evaluation and documentation interface cable AK
540/B (902842) and the PC communication program
Achat OC (208990) are required. |
|
 |
208205 OxiTop® Control 12
BOD measuring instrument for self-check, with
controller OC100, 6 measuring units with IR
interface, 115 VAC, 50/60 Hz, and accessories. For
PC evaluation and documentation interface cable AK
540/B (902842) and the PC communication program
Achat OC (208990) are required. |
|
|
|
BOD
INSTRUMENTATION ACCESORIES |
208990 Achat OC
PC communication program for Controller OxiTop® OC
100 for further data processing and documentation of
measuring data. Required accessories: Interface
cable AK 540/B (902842). |
|
|
902842 AK 540/B
Interface cable for connection of instrument with
6-pin socket to a PC via RS 232 interface, 9/25
adapter included. |
|
|
208433 OxiTop® Thermostat Box
OxiTop® BOD Instrumentation thermostat box, for up
to 12 OxiTop® units and stirrer base, 115 VAC, 50 Hz
|
|
|
|
RELATED
PRODUCTS |
WS700
Composite/Discrete Water Sampler
Single pump composite or discrete water sampler.
|
|
|
WS750
Stormwater/Wastewater Sampler
Dual pump stormwater or wastewater sampler.
|
|
|
|
WS700R
Refrigerated Composite Sampler |
|
|
FSS-STD Flow Sampling System
A unique combination of instruments that makes
monitoring, sampling, and recording stormwater or
industrial discharge water flows easy. |
|
|
WQS-STD Water Quality Sampling System
A unique combination of instruments that allows for
monitoring, sampling, and recording of water flows
that exceed a water quality parameter such as
turbidity, pH, conductivity, ect. |
|
 |
|
3 FT Long Handled Dipper |
|
|
|
Swing Sampler w/Pole & Bottle |
|