Permanent Instrument
ACTEON 2053-TU SC-T
New self-cleaning turbidity and temperature field transmitter

 

  • Sturdy, watertight, easy to install

  • Widescreen graphic display: instant measurements, trend line, calibration state

  • Quick and simple intuitive programming

  • 4-20 mA outputs, programmable relays

 

 

Technical Specifications
Ranges 0.0-200.0 or 0-2000 NTU / 0.0-200 or 0-2000mg/l
Measurement accuracy ±1 NTU
Temperature measurement range -10.00 to +50.00°C
Temperature measurement accuracy (°C) ±0.1°C
   
Casing ABS
Protection IP 65
Operating temperature -25°C to +55°C
Dimensions (L x W x D) / Weight 173 z 195 x 103mm / 1.5kg
Display Widescreen back-lit graphic: 240 x 128 pixels (108 x 58mm)
   
Power supply 230/115 VAC 60Hz, Optional: 24Vdc
Max power consumption 10VA
   
4-20mA outputs

2 galvanic isolation outputs (max load 700ohms):

- Adjustable from 0.0 to 200.0 NTU or from 0 to 2000 NTU

- Adjustable from -10.00°C to +50.00°C

Relay outputs

2 relays that can be configured in different modes:

-Adjustment in alarm mode (1 TU and 1 temperature (°C) threshold)

-Adjustment in adjustment mode (2 TU thresholds)

 

Fields of application:

  • Turbidity is a qualitative reference parameter for the residual charge of undissolved substances.

  • Turbidity monitoring in wastewater treatment structures help to determine if these purification systems are efficient:

  • Wastewater treatment (input, treatment line, sludge, output)

Simple, sturdy and reliable like its legendary predecessor, the APF series with additional digital intelligence and leading-edge technology.

 

The optics are cleaned by a fast-action scraper. This is a polyurthane scrapper powered by a stainless steel piston. When the scraper is at rest, it is completely embedded in its housing. This piston system (Ponsel invention) prevents fibres from getting entwined around the axle. A control unit (SNA) can be used to program how often the cleaning is to be carried out: 5 strokes every 15, 30, 60 minutes.

 

SNA unit: Electronic box for controlling the turbidity probe's self-cleaning system.

Dimensions: (L x W x D): 120 x 160 x 130 mm
Weight: 0.420kg
Materials: Reinforced polyester with transparent
Cover: IP 65

 

The ACTEON 2053 unit comes with a PONCIR-TU20-NA-10 optical infrared sensor.

 

Self-cleaning turbidity sensor: PONCIR-TU20-NA-10:

 

Measurement principle: Optical absorption of infrared light, Regulated and pulsed emission (10Hz frequency)
Dimensions: 253 x 73.2mm
Weight: 1kg
Material: PVC
Temperature compensation (C): Automatic, by NTC thermistor from +50 to +30°C
Emission wavelength: 880mm (infrared)
Cable: Multiple coated wires, polyurethane sheath. 10m standard length         (up to 100m on request)
Installation options: Fixed sensor-holder perch for in-pipe installation.

 

-Modulation electronics system: Preamplification and optical feedback control integrated into the shaft of the nominal power supply sensor: Unregulated common analog voltage (6 to 12V), consumption less than 2mA under 6V.

 

Temperature Sensor:
Delrin, Stainless steel, Silicone - IP68
Dimensions: Diameter 3mm (sensor), Length 115mm
Weight: 40g
Cable: 10m standard length

 

Accessories for electronic unit installation:

 

 


  Back to Ponsel Main Page    |  M2M Telemetry/SCADA HMI    |   Back to Testech Home Page